TiControl - Inspiratory time control
S, ST and iVAPS modes only
Unique to ResMed bilevel devices, TiControl
allows the clinician to set minimum and maximum
limits on the time the device spends in IPAP. The minimum and maximum time limits are set at
either side of the patient’s ideal spontaneous inspiratory time, providing a ’window of opportunity’
for the patient to spontaneously cycle to EPAP.
The minimum time limit is set via the Ti Min parameter and the maximum time limit is set via the Ti
Max parameter.
TiControl’s Ti Max and Ti Min parameters play a significant role in maximizing synchronization by
effectively intervening to limit or prolong the inspiratory time when required. This ensures
synchronization, even in the presence of significant mouth and/or mask leak.
The following table is a guide to selecting the Ti Max and Ti Min values that best correspond to the
patient’s respiratory rate and inspiration and expiration ratio, depending on the respiratory
conditions.
Patient breath
(BPM)
Ttot = 60/BPM
(sec)
I:E = 1:2
(Reference)
Sufficient inhalation time
I:E = 1:1
Secure
exhalation time
I:E = 1:3
Ti Min
Ti Max
Ti Max
10 6 2 1.0 2.0 1.5
15 4 1.3 1.0 2.0 1.3
20 3 1.0 0.8 1.5 1.0
25 2.4 0.8 0.7 1.2 0.8
30 2 0.7 0.6 1.0 0.7
35 1.7 0.6 0.5 0.8 0.7
40 1.5 0.5 0.5 0.7 0.7

Table of Contents

Other manuals for ResMed AirCurve 10 ST-A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirCurve 10 ST-A and is the answer not in the manual?

ResMed AirCurve 10 ST-A Specifications

General IconGeneral
BrandResMed
ModelAirCurve 10 ST-A
CategoryMedical Equipment
LanguageEnglish

Related product manuals