10
HOW TO ATTACH A HUMIDIFIER
ATTACH A HUMIDAIRE 3
i
HUMIDIFIER
Please refer to the illustration in section E of the illustration sheet, and your HumidAire 3i
manual.
The
HUMIDAIRE 3
i
attaches to the front of the S8 COMPACT to provide heated
humidification. No other accessories are required for its use.
Notes:
a. Make sure the water chamber of the HumidAire 3i is empty before you attach or
detach the humidifier.
b. Remove the connector plug on the S8 Compact (see section
A of the illustrations
sheet) before you attach the HumidAire 3i. Reattach the connector plug firmly
after you detach the HumidAire 3i.
ATTACH A HUMIDAIRE HUMIDIFIER
Please refer to the illustrations in section F of the illustration sheet.
To connect your
S8 COMPACT to a HUMIDAIRE humidifier you need medium sized
21” (52 cm) air tubing.
1 Fill the HUMIDAIRE water chamber and place inside the HUMIDAIRE. Connect the
medium 21” (52 cm) and long 6‘6” (2 m) air tubing (F-1), and close the humidifier lid.
2 Place the S8 COMPACT on top of the HUMIDAIRE (F-2) and connect the free end of
the medium air tubing to the air outlet (F-3).
Note: To avoid water spilling into the S8 Compact, do not put it underneath the
humidifier.
3 Assemble your mask system and connect it to the free end of the long air tubing (F-4).
4 Plug the HUMIDAIRE power cord into a power outlet. Connect the S8 COMPACT power
cord to the socket at the rear of the
S8 COMPACT and plug the other end into a power
outlet.
Please refer to the HumidAire User’s Manual for further details.
!
WARNING
Make sure that the power cord and plug are in good condition and the equipment
is not damaged.
The
S8 COMPACT is now ready for use with the HUMIDAIRE.