EasyManuals Logo

ResMed S8 Compact User Manual

ResMed S8 Compact
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
20
TROUBLESHOOTING
If there is a problem, try the following suggestions. If the problem cannot be solved,
contact your equipment supplier or ResMed. Do not attempt to open the device.
Problem Possible Cause Solution
No display. Power not connected. Ensure the appropriate power
cord is connected to your
S8 Compact and the power
switch (if available) is on.
Insufficient air
delivered from
S8 Compact.
Ramp Time is in use. Wait for air pressure to build up
or change ramp time.
Air filter is dirty. Replace air filter.
Air tubing is kinked or
punctured.
Straighten or replace tubing.
Air tubing not connected
properly.
Check air tubing.
Mask and headgear not
positioned correctly.
Adjust position of mask and
headgear.
Cap(s) missing from access
port(s) on mask.
Replace cap(s).
Altitude compensation setting is
incorrect.
Adjust altitude compensation
setting to the correct altitude
range.
Pressure required for treatment
may have changed.
See your clinician to adjust the
pressure.
338249r1.book Page 20 Friday, June 10, 2005 3:22 PM

Table of Contents

Other manuals for ResMed S8 Compact

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed S8 Compact and is the answer not in the manual?

ResMed S8 Compact Specifications

General IconGeneral
BrandResMed
ModelS8 Compact
CategoryMedical Equipment
LanguageEnglish

Related product manuals