27
SYSTEM SPECIFICATIONS
ENGLISH
Recommended separation distances between portable and mobile RF communications
equipment and the S8 series of devices
The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled.
The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the S8 device
as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output
power of transmitter
W
150kHz to 80MHz
d = 1.17 P
80 MHz to 800 MHz
d = 0.35 P
800MHz to 2.5 GHz
d = 0.7 P
0.01 0.17 0.04 0.07
0.1 0.37 0.11 0.22
1 1.17 0.35 0.7
10 3.69 1.11 2.21
100 11.70 3.50 7.0
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
338249r1.book Page 27 Friday, June 10, 2005 3:22 PM

Table of Contents

Other manuals for ResMed S8 Compact

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed S8 Compact and is the answer not in the manual?

ResMed S8 Compact Specifications

General IconGeneral
BrandResMed
ModelS8 Compact
CategoryMedical Equipment
LanguageEnglish

Related product manuals