EasyManua.ls Logo

Respironics REMstar Plus with C-Flex - Device Specifications; Technical Specifications

Respironics REMstar Plus with C-Flex
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 ❍ REMstar Plus with C-Flex User Manual
Specifications
Dimensions: 9 1/2” x 6 7/8” x 4 1/2” (24 x 17 x 12 cm)
Weight: <4.0 lbs. (<1.8 kg)
Operation Transport & Storage
Temperature: 41 to 95° F (5 to 35° C) -4 to 140° F (-20 to 60° C)
Humidity: 15 to 95%
Non-condensing
15 to 95%
Non-condensing
Atmospheric Pressure: 76.7 to 102 kPascals N/A
This device is designed to conform to the following standards:
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment
IEC/EN 60601-1-2:2001 (2nd Edition) Immunity for Medical Electrical Equipment
EN ISO 17510-1 Sleep Apnea Breathing Therapy Devices
Continuous
100 - 240 VAC, 50/60 Hz, 1.0 A max.
12 VDC, 3.0 A max.
Class II Equipment
Type BF Applied Part
IPX0 - Ordinary Equipment
4 to 20 cm H
2
O (in 1 cm H
2
O increments)
4 to 20 cm H
2
O (±1.0 cm H
2
O)
Measured in accordance with EN ISO 17510 @ 6.6, 13.2, & 20 cm H
2
O @ 500 ml with BPM set to 10, 15, & 20
BPM performed at 23° C (±2° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
<30 dB(A)
Measured in accordance with EN ISO 17510 @ 10 cm H
2
O at the patient circuit. This measurement
applies to the CPAP device with or without the optional REMstar Heated Humidifier.
35 LPM
Measured in accordance with EN ISO 17510 @ 6.6, 13.2, & 20 cm H
2
O @ 500 ml with BPM set to 10, 15, & 20
BPM performed at 23° C (±2° C), 50% RH (±5%), and an atmospheric pressure of 101.54 kPascals.
±(0.15 + 4% of actual reading)
Device Size
Product Use,
Transport, and
Storage
Standards
Compliance
AC Power Consumption
Type of Protection
Against Electric Shock
Degree of Protection
Against Electric Shock
Degree of Protection
Against Ingress of Water
Pressure Range
Mode of Operation
Pressure Stability
Sound Pressure Level
Maximum Flow
DC Power Consumption
Pressure Display Accy.