Chapter 8: Testing
Tranquility Quest Plus CPAP System Service Manual
29
1003740
8.2 Tranquility Quest Plus System Final Test Data Sheet
Pre Run In Time and Date Post Run In Time and Date Total Run In Time
Serial Number Service Notification (RI use)
Model Number
Test Step Specification Actual Actual Final
Pre Run-In Post Run-In Pass / Fail
LED Illuminated 2 Pass / Fail
Minimum Pressure 2.5 to 3.5
Occluded 3 cm H
2
O
Minimum Pressure 2.0 to 4.0
Unoccluded 4 cm H
2
O
Maximum Pressure 6 19.0 to 21.0
cm H
2
O
Ramp 7 Pass / Fail
Pressure after Run - In 9 9.0 to 11.0
cm H
2
O