Chapter 2: Specifications
Tranquility Quest Plus CPAP System Service Manual
4
1003740
ELECTRICAL:
AC Input Voltage................................... 85 - 260VAC; 47-63Hz
AC Current Consumption .................... 1.0 A maximum
Power Consumption ............................. 75 VA maximum
Class......................................................... IIa
Type ......................................................... Type BF
RAMP:
Minimum Ramp Pressure............. 3.0 - 20.0 cm H
2
O
Ramp Resolution.............................. 1.0 cm H
2
O
Ramp Time ....................................... 0 - 30 minutes
Ramp Time Resolution ................... 5 minute increments
PRESSURE:
CPAP Pressure ....................................... 3.0 – 20.0 cm H
2
O
Readout Resolution ............................... 1.0 cm H
2
0
Accuracy, Static ...................................... +1.0 cm H
2
O for flows between
-60 to + 120 L/min.
Accuracy, Dynamic ............................... + 1.5 cm H
2
O for flows between
-60 to + 120 L/min.
Chapter 2: Specifications
2.1 Specifications
ENVIRONMENTAL:
Operating Temperature ........................ 41 to 104°F (5 to 40°C)
Storage Temperature............................. -4 to 140°F (-25 to 60°C)
Humidity ................................................ 10 to 95% non-condensing
Altitude ................................................... 0 - 10,000 ft (0 - 3048 m)
Noise Level ............................................. No specification is given
because various test instru-
ments, test procedures, and unit
operating conditions produce
varying results.
PHYSICAL:
Dimensions ............................................. 9.5” (L) x 7.75” (W) x 5.0”(H)
(24.1 x 19.7 x 5.0cm)
Weight ..................................................... 4.9 lbs (2.2 kg)
FUSE:
Fuse Type ................................................ T1A, 250v, 5 x 20 mm
(Respironics P/N 1004092)