Chapter 8: Testing
Tranquility Quest Plus CPAP System Service Manual
31
1003740
8.4 Tranquility Quest Plus System Final Test Data Sheet
Clinical Remote Control (CRC)
(Optional to be used if CPAP unit is equipped with a CRC)
Serial Number Service Notification (RI use)
Model Number
Test Step Specification Actual Actual Final
Pre Run-In Post Run-In Pass / Fail
CRC Setting 17.7 to 18.3
18 cm H
2
O cm H
2
O
CRC Setting 14.7 to 15.3
15 cm H
2
O cm H
2
O
CRC Setting 8.7 to 9.3
9 cm H
2
O cm H
2
O
CRC Setting 5.7 to 6.3
6 cm H
2
O cm H
2
O
CRC Setting 2.7 to 3.3
3 cm H
2
O cm H
2
O
CRC Ramp LED
Illuminated and Ramp Pass / Fail
Operation
Tested By: (Print) Tested By: (Signature) Date