EasyManua.ls Logo

RF-KIT RF2K-S - Adjustment Forward Power Display; Fine Adjustment Power Display Via the Menu

RF-KIT RF2K-S
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Assembly and adjustment instructions RF-KIT Power Amplifier RF2K-S
Step 19: Severe Error Alarm
Connect TRX to Exciter input (17).
Connect Dummy load via Reference wattmeter to ANT1 (6).
Tune TRX to a frequency in the 20 m band (14 MHz).
Adjust R19 on the Controller board about 10 turns counterclockwise.
Switch the device to “Operate“ by tapping switching area Standby.
Set the transmit power of the TRX to 5 W (CW/AM/FM) reading on your
reference wattmeter.
Operate PTT.
If a Severe Error Alarm is already triggered now, R19 on the
Controller board must be adjusted further counterclockwi-
se. This will continue until the alarm message disappears.
Gradually increase the transmit power of the TRX until 500 W is dis-
played on the reference wattmeter (the power display of the PA
display may still show a different value, as it will be calibrated in the
next step).
Keep PTT pressed.
Adjust the PAF display with R19 on the Controller board until the
value “1100“ is displayed.
(Menu Calibration Poti Config: “Indication area PAF“)
Release PTT, stop transmitting.
Step 20: Adjustment forward power display (full power)
This adjustment is carried out in the 80 m-Band (3.5 MHz).
Connect TRX to Exciter input (17).
Connect Dummy load via Reference wattmeter to ANT1 (6).
Operate PTT.
Slowly increase the TRX transmit power from 5 W until Reference
wattmeter reads 1000 W (CW/AM/FM).
Set R24 on the Controller board so that on the Touch screen (2)
(main screen) Forward power is displayed with 1000 W.
Release PTT, stop transmitting.
Step 21: Fine adjustment power display via the menu
Please also note the description in the operating instructions
(page 14, “Oset Calibration“)
Go to
Menu Calibration Oset Calibration:
The unavoidable frequency-dependent deviations of the linearity of the
directional coupler used for internal power measurement can be minimi-
zed for each bandpass range. The adjustment should be performed at an
output power of 1 kW. Use your preferred reference wattmeter for the ad-
justment.
Selection of the bandpass range is done automatically during transmission.
The detected band is highlighted.
Adjust the internal power display by tapping switching areas
> (increase value) respectively < (decrease value).
Brief actuation changes the value in the indication area Forward by 0.1 W.
Longer actuation changes the value continuously.
Press the switching area Save to store the settings.
Repeat this procedure for each bandpass range to be adjusted.

Other manuals for RF-KIT RF2K-S