RIGOL Chapter 6 MATH and Measurements
6-20 MSO2000A/DS2000A Users Guide
Note:
Negative delay value/negative phase value indicates that the edge selected of
source A occurred after the edge selected of source B.
Phase is expressed in degree and the calculation formula is as shown below.
°360×=
PeriodA
Delay
Phase
Wherein,
Phase
is the phase measurement result;
Delay
is the delay
measurement result;
PeriodA
is the period of source A.