RIGOL DG1000 Command System
Programming Guide for DG1000
2-56
MEMory
The generator has 10 storage locations (STATE1 to STATE10) in non-volatile memory
to store instrument states. The locations are numbered from 1 to 10. Location 0 is
volatile memory and the generator automatically uses location “0” to hold the state of
the instrument at power-down. MEMory commands can be used to read the location
names, delete the stored instrument states, automatically recall instrument state and
query the available storage locations. You can also assign a user-defined name to
each of the locations (1 to 10) from the front panel.
DG1000 supports following MEMory commands:
1. MEMory:STATe:NAME
2. MEMory:STATe:NAME?
3. MEMory:STATe:DELete
4. MEMory:STATe:RECall:AUTO
5. MEMory:STATe:RECall:AUTO?
6. MEMory:STATe:VALid?
7. MEMory:NSTates?