RIGOL Chapter 2 Calibration Process
2-2 DP700 Calibration Guide
Figure 2-2 Calibration Password Input Box
(4) Press OK to confirm your password, and then you enter the calibration
page, as shown in Figure 2-3.
Figure 2-3 Calibration Page
Step 2 Select the calibration item
In the calibration page, press Up/Down arrow key or use the knob to switch the
calibration items. The calibration items for DP700 series are available in Table
2-1. You can select one or multiple items for calibration. RIGOL recommends
that you perform the calibration for all the available items.
Table 2-1 Calibration Items
Calibration Item
V DAC (calibrate the accuracy of the output voltage)
2
I DAC (calibrate the accuracy of the output current)
V ADC (calibrate the accuracy of the readback voltage)
4
I ADC (calibrate the accuracy of the readback current)
OVP (calibrate the accuracy of the overvoltage protection setting value)