Command Systems RIGOL
Programming Guide for DS1000B Series
2-67
:MEAS:DELASOUR? Return CH1, CH2.
27. :MEASure:PHAseSOURce
Command Format:
:MEASure:PHAseSOURce <source>,<source>
:MEASure:PHAseSOURce?
Functions
The command selects the channel for measuring the phase delay. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns CH1, CH2 or CH1, CH3, CH1, CH4, CH2, CH3, CH2, CH4, CH3,
CH4.
Example:
:MEAS:PHASOUR CHAN1 CHAN2 Measure the phase delay.
:MEAS:PHASOUR? Return CH1, CH2.
28. :MEASure:ENABle
Command Format:
:MEASure:ENAble
Functions
This command is open the locked MEASURE button and allow user to open Auto
Measure.
NOTE: Only corresponding unlock command can open AUTO and Auto Measure
once they are locked, neither restart nor update is available.
29. :MEASure:DISable
Command Format:
:MEASure:DISable
Function: