RIGOL Command Systems
Programming Guide for DS1000B Series
Yor: vertical offset relative to YREF;
Yref: Y reference, the middle point of the screen.
Example:
+1,+0,0,+1,8.000e-009,-6.000e-006,+0,4.000e-002,0.000e000,+100
7. :WAVeform:YINCrement?
Command Format:
:WAVeform:YINCrement? [<source>]
Function:
This command queries the Y unit voltage of the specified source. <source> can
be: CHANnel1, CHANnel2, CHANnel3, CHANnel4 or MATH.
NOTE: returned value= VoltScale /25
Returned Format:
The query returns Y unit voltage, and the unit is V.
Example:
:WAV:YINC? CHAN2 Return 4.000e000.
8. :WAVeform:YORigin?
Command Format:
:WAVeform:YORigin? [<source>]
Function:
The command queries the vertical offset of specified source. The <source> may
be CHANnel1, CHANnel2, CHANnel3, CHANnel4 or MATH.
Returned Format:
The query returns the value of vertical offset, and the unit is V.
Example:
:WAV:YOR? CHAN2 Return -1.600e001.