Chapter 1 Disassemble and Assemble RIGOL
MSO1000Z/DS1000Z Service Guide 1-13
Assemble Procedures
The assemble procedures are the reverse of the disassemble procedures. Check
whether the cables are correctly connected and whether all the screws are installed
after each step of assemble.
You are recommended to follow the order and methods introduced above
when disassembling and assembling the instrument to avoid damage to
the instrument due to improper operation and to save your time.