Understand the Horizontal System   
The oscilloscope shows the time per division in the scale readout. Since all active 
waveforms use the same time base, the oscilloscope only displays one value for all 
the active channels, except when using Delayed Scan, or Alternative Trigger. 
The horizontal controls changes the horizontal scale and position of waveforms. The 
horizontal center of the screen is the time reference for waveforms. Changing the 
horizontal scale causes the waveform to expand or contract with respect to the 
screen center. 
Horizontal position changes the displayed waveform position, relative to the trigger 
point. 
 
The Horizontal Knobs 
: The horizontal    knob adjusts the horizontal position of 
all channel (include Math) waveforms. The resolution of this control 
varies with the time base. Pressing this button clears trigger offset 
and moves the trigger point to the horizontal center of the screen. 
: Use   to select the horizontal time/div (scale factor) for the 
main or the Delayed Scan time base. When Delayed Scan is enabled, it 
changes the width of the window zone by changing the Delayed Scan 
time base. 
 
Horizontal Menu 
Press the horizontal MENU button to display the horizontal menu. The settings of this 
menu are listed in the following table.