Example 5: Making Cursor Measurements
There are 20 build-in automatic measurements. They can also be conducted using
cursors to make time and voltage measurements of a waveform quickly.
Measure the Peak Frequency of the First Sinc Waveform
To measure the ringing frequency at the rising edge of a signal, do these steps:
1. Press Cursor key to see the Cursor menu.
2. Press Mode to set Manual mode.
3. Press Type to select X.
4. Turn (
) knob to place cursor A on the first peak of the wave.
5. Turn (
) knob to place cursor B on the second peak of the wave.
Observe the delta in time and frequency displayed on the screen.
Figure 3-4
Waveform display