RIGOL 6 To Make Measurements
6-20 DS6000 User’s Guide
threshold middle values the falling edge of source A and the rising edge of
source B and can be calculated according to "Delay A
B " and the period of
source A.
The phase calculation formula:
°×= 360
PeriodA
Delay
Phase
Wherein,
Phase
represents Phase,
Delay
represents Delay,
PeriodA
represents the period of source A.
Note: The default value of the threshold middle value is 50%. You can also set it via
MEAS Setting Type "Threshold". For the setting method, refer to
"Measurement Setting".