6. User Mode Settings
6-4. Bump test expiration setting (BUMP
.SET)
- 77 -
3
Press the
AIR button several times to select the
behavior after expir
ation, then press the
POWER/MODE button.
Select [CONFIRM], [CANT
.USE], or [NONE].
[END] appears once the settings are complete. The
display returns to the [BP
.EXPD] screen.
76
78
Table of Contents
Default Chapter
2
Table of Contents
2
1 Product Overview
5
Introduction
5
Intended Use
7
DANGER, WARNING, CAUTION, and NOTE
8
2 Important Safety Information
9
Danger Information
9
Warning Information
12
Caution Information
15
Safety Information
19
3 Product Configuration
21
Main Unit and Accessories
21
Part Names and Functions
23
Main Unit
23
LCD Display
25
Inserting the Batteries
26
4 Alarm Functions
30
Gas Alarm Types and Alarm Setpoints
30
Gas Alarm Activation
34
Fault Alarm Activation
36
Outside Operating Temperature Range Warning
37
5 Usage Instrucions
38
Usage Note
38
Preparing Startup
38
Turning on the Power
39
Performing Air Calibration
45
Measuring Gas Concentration
48
Checking the Gas Concentration, Alarm Setpoints, Etc. (Display Mode)
50
Procedure for Displaying Display Mode
50
Items Displayed in Display Mode
51
Turning off the Power
56
6 User Mode Settings
57
User Mode Setting Procedure
57
User Mode Setting Items
60
Calibration Expiration Setting (CAL SET)
64
Calibration Expiration ON/OFF (CAL.RMDR)
65
Calibration Expiration Date Interval (CAL.INT)
66
Setting Operation after Calibration Date Expires (CAL.EXPD)
67
Bump Test Expiration Setting (BUMP.SET)
69
Bump Test Setting (SETTING)
70
Bump Test Expiration ON/OFF (BP.RMDR)
74
Bump Test Expiration Date Interval Setting (BP.INT)
75
Setting Behavior after Bump Test Expiration (BP.EXPD)
76
Alarm Setpoint Setting (ALARM-P)
78
Alarm Setpoint Setting (ALARM-P)
79
Resetting Alarm Setpoints (DEF.ALMP)
85
Lunch Break ON/OFF (LUNCH)
86
Confirmation Beep Setting (BEEP)
87
Beep Operation Setting (BEEP.SEL)
88
Beep Interval Setting (BEEP.INT)
89
LCD Lighting Time Setting (BL TIME)
90
Key Operation Tone ON/OFF (KEY.TONE)
91
Display Mode Item Display ON/OFF (DISP.SET)
92
Quick Calibration Time Setting (E-CAL)
93
Date and Time Setting (DATE)
94
User Mode Password Setting (PASS-W)
95
ROM/SUM Display (ROM/SUM)
96
7 Maintenance
97
Maintenance Intervals and Maintenance Items
97
Performing Calibration
99
Preparation for Calibration
99
Displaying the Calibration (GAS CAL) Screen
108
Performing Air Calibration
109
Performing AUTO Calibration
112
Performing Bump Tests
121
Performing Alarm Tests
127
Cleaning Instructions
129
Parts Replacement
130
Periodic Replacement Parts
130
Filter Replacement
132
Sensor Replacement
136
8 Storage and Disposal
140
Procedures for Storage or When Not in Use for Extended Periods
140
Product Disposal
141
9 Troubleshooting
142
Product Abnormalities
142
10 Product Specifications
144
Common Specifications
144
Specifications by Model
146
11 Appendix
148
Data Logger Function
148
Terminology
151
Other manuals for Riken Keiki 04 Series
Operation Manual
81 pages
User Guide
2 pages
4
Based on 1 rating
Ask a question
Give review
Questions and Answers:
Need help?
Do you have a question about the Riken Keiki 04 Series and is the answer not in the manual?
Ask a question
Riken Keiki 04 Series Specifications
General
Brand
Riken Keiki
Model
04 Series
Category
Measuring Instruments
Language
English
Related product manuals
Riken Keiki GX-8000
46 pages
Riken Keiki GX-3R Pro
112 pages
Riken Keiki RX-415
33 pages
Riken Keiki FI-8000
59 pages