7. Maintenance 7-2. Performing calibration
- 99 -
7-2. Performing calibration
The product can be calibrated using AUTO calibration with preset gas concentrations, in addition to air calibration
performed in the atmosphere.
Calibration requires dedicated tools and a calibration gas. Contact Riken Keiki before performing calibration.
7-2-1. Preparation for calibration
Prepare the following equipment and calibration gases, then connect to this product:
<Required equipment>
Gas set
(Gas sampling bag, tube (length not exceeding 1 m), pump with flow adjuster function (a pump, flow gauge, and
needle can also be used.))
Calibration adapter
Stopwatch
<Calibration gases and recommended gas concentrations>
The calibration gases and recommended gas concentrations are as follows. (Domestic standard)
Model Calibration gas Recommended value
OX-04G, OX-04 N
2
99.9 % or more
CO-04 CO (N
2
-based) 80 ppm
HS-04 H
2
S (N
2
-based) 16 ppm
CO-04 (C-)
CO (N
2
-based) 80 ppm
H
2
(air-based)
*
500 ppm
*H
2
(air-based) must be calibrated at ambient temperatures ranging from 10 °C to 30 °C.

Table of Contents

Other manuals for Riken Keiki 04 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Riken Keiki 04 Series and is the answer not in the manual?

Riken Keiki 04 Series Specifications

General IconGeneral
BrandRiken Keiki
Model04 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals