6. User Mode Settings 6-4. Bump test expiration setting (BUMP.SET)
- 76 -
6-4-4. Setting behavior after bump test expiration (BP.EXPD)
This selects the behavior after the bump test expiration date.
Select one of the following:
CONFIRM: The behavior differs depending on the operation.
Press the AIR button to proceed to measurement mode or press the POWER/MODE button to
proceed to the bump test.
CANT.USE: Measurement mode is not available. Press the POWER/MODE button to proceed to the bump test.
If you do nothing, the product will automatically proceed to the bump test after about six seconds.
NONE: The behavior differs depending on the operation.
If the product indicates that the expiration date has been reached, press the POWER/MODE button
to proceed to the bump test. If you do nothing, the product will proceed to measurement mode after
about six seconds.
The default setting is CONFIRM.
1 Press the AIR button several times in user
mode to select [BUMP.SET], then press the
POWER/MODE button.
2 Press the AIR button several times to select
[BP.EXPD], then press the POWER/MODE
button.