EasyManua.ls Logo

RITEC RPR-4000 - Page 20

RITEC RPR-4000
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RPR-4000
Operations manual
REV 8-21 20
6. View the broadband receiver output on the oscilloscope. You should see the
signal from your signal generator amplified to 1 V peak to peak.
7. Please note that the receiver has a blanking function which will not be coherent
with your signal. Because of this seen you may see a randomly appearing or
moving blank space in your signal on the oscilloscope. This is normal.
Setting up a pulse-echo measurement
Operating the RPR-4000 Pulser/Receiver in the two standard modes of operation (Pulse-
Echo and Through Transmission) is very straight forward.
For Pulse-Echo operation, proceed as follows:
1. Connect a short coaxial cable from “Diplexer Output to Receiver” to Receiver
“Input Number 1”.
2. If using a piezo-electric transducer, connect coax cables from “High Power RF
Pulse Output” to an RT-150 (150 ohm high power load) and then to an
appropriate transducer.
3. If a low impedance transducer (e.g., an electromagnetic acoustic transducer) is
used, then 150 ohm load is not required.
4
4. Connect a coaxial cable from “Trigger Out” on the rear panel of the RPR-4000 to
the external trigger input on the oscilloscope.
5. Connect a coaxial cable from receiver “Output” to a 50 ohm vertical input on the
scope. Set scope vertical sensitivity to 0.5 V per division.
6. Connect a coaxial cable from “RF Pulse Monitor” to a 50 Ω terminated input on
the oscilloscope. Set the vertical sensitivity to 1 V per division.
(Because the monitor output is down 40 dB [100:1] this vertical sensitivity then
corresponds to 100 V at the pulser output per division).
7. Using the keypad on the front panel set the various operating parameters as
follows:
Page 1 – Pulser settings 1/2
Frequency: Transducer center frequency
Pulse Width: 5 cycles
4
The 150 Ω termination on the output serves two functions: 1.) Cleans up the base line
during the RF pulse and 2.) Allows the control algorithm for output level to function with
higher resolution.