EAGLE 2 Operator’s Manual PID Calibration • 145
5. The Calibration Mode Screen displays with the cursor next to AUTO
CALIBRATION.
6. Move the cursor to the PERFORM AIR ADJUST menu item by using
the RANGE SHIFT button.
7. Press and release the POWER ENTER RESET button. The following
screen appears.
8. Press and release the AIR YES button to continue.
If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.
CALIBRATION MODE
> AUTO CALIBRATION
SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION
PERFORM
AIR ADJUST?