72 • Replacing or Recharging the Batteries EAGLE 2 Operators Manual
Replacing or Recharging the
Batteries
WARNING: To prevent ignition of a hazardous atmosphere, batteries
must only be changed or charged in an area known to be
nonhazardous.
Replace or charge the batteries when the EAGLE 2 indicates that it is in low
battery warning. When in low battery warning, BATT appears vertically
along the left side the LCD.
Replacing the Batteries
NOTE: Use Duracell Procell PC 1400 alkaline batteries, Duracell MN 1400
alkaline batteries, Energizer E93 or Energizer EN93, or RKI
Instruments, Inc. 49-1330RK Ni-MH batteries to maintain the CSA
classification of the EAGLE 2. Use of other batteries or mixing
alkaline and rechargeable batteries will void the CSA classification
and may void the warranty.
1. Turn off the EAGLE 2.
WARNING: Do not remove the batteries while the EAGLE 2 is on.
2. Loosen the battery case thumbscrew by turning it counterclockwise until
it disengages from the bottom case. If necessary, use a coin or large flat
blade screwdriver to loosen it.
CH4 0%LEL
B OXY 20.9vol%
A H2S 0.0ppm
T CO 0ppm
T