EAGLE 2 Operator’s Manual Infrared Hydrocarbon Calibration • 257
5. The Calibration Mode Screen displays with the cursor next to AUTO
CALIBRATION.
6. Move the cursor to the PERFORM AIR ADJUST menu item by using
the RANGE â–¼ SHIFT button.
7. Press and release the POWER ENTER RESET button. The following
screen appears.
8. Press and release the AIR â–² YES button to continue.
If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.
CALIBRATION MODE
> AUTO CALIBRATION
SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION
PERFORM
AIR ADJUST?