EasyManua.ls Logo

RKI Instruments EAGLE 2 - Page 282

RKI Instruments EAGLE 2
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
272 Start Up, Bar Hole Mode EAGLE 2 Operator’s Manual
NORMAL MODE and LEAK CHECK MODE will be the
only choices displayed. In this case, see “Turning On the
EAGLE 2” on page 22 or “Turning On the EAGLE 2, Leak
Check Mode” on page 280.
1. Connect the sample hose to the EAGLE 2’s quick connect inlet
fitting.
2. Connect the bar hole probe to the sample hose’s quick connect
fitting.
3. Press and briefly hold down the POWER ENTER RESET button.
Release the button when you hear a beep.
4. The LCD will show the following screen for about ten seconds.
5. The Battery Voltage Screen appears for a few seconds.
6. The Mode Select Screen displays.
7. The cursor will flash in front of NORMAL MODE. Use the
RANGE SHIFT button to move the cursor next to BAR HOLE
MODE.
8. With Bar Hole Mode selected, press and release the POWER
ENTER RESET button to begin the Bar Hole Mode startup
sequence.
GAS MONITOR
EAGLE 2
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
> NORMAL MODE
BAR HOLE MODE
LEAK CHECK MODE

Table of Contents

Other manuals for RKI Instruments EAGLE 2

Related product manuals