EAGLE 2 Operator’s Manual Display Mode • 45
The following is the complete list of factory defined gases.
2. Use the AIR ▲ YES or RANGE
▼ SHIFT buttons to move the cursor
next to the desired gas.
3. Press and release POWER ENTER RESET. The catalytic combustible
channel will be configured to the selected gas and the EAGLE 2 will
proceed to the STEL Screen. This configuration will be in force until
either a different gas is selected in Display Mode or the unit is turned off.
NOTE: If a PID sensor is installed in the EAGLE 2 and CATALYTIC
SENSOR RELATIVE RESPONSE is set to ON in Setup Mode, a
PID Sensor Relative Response Screen appears after the Catalytic
Sensor Relative Response Screen. If CATALYTIC SENSOR
RELATIVE RESPONSE is set to OFF, the PID Sensor Relative
Response Screen will still appear after the Methane Elimination
Mode Screen. See “PID Relative Response Feature” on page 141 for
a description of the relative response feature for the PID sensor.
NOTE: If Methane Elimination Mode is enabled, the Catalytic Sensor
Relative Response screen does not appear.
Table 7: Relative Response Gas List
Acetone Cumene Isobutane Methyl Ibutyl
Ketone
Mixed Xylenes
Benzene Ethylene
Dichloride
Isopropanol Nonane Field Defined Gas
Butyl Acrylate Ethyl Alcohol Methane Pentane Field Defined Gas
Butyl Acetate Ethyl Chloride Methanol Propane Field Defined Gas
2-Butyl Alcohol Ethyl Acrylate Methyl Acetate Styrene Field Defined Gas
1-Butyl Alcohol Hexane Methyl Acrylate Toluene Field Defined Gas
Cyclohexane Hydrogen Methyl Ketone Vinyl Acetate