EasyManua.ls Logo

RMF Systems CMS 2 - Test Status; Other Faults

Default Icon
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
As a policy of continual improvement, RMF reserves the right to alter the specification without prior notice.
201.028 REV 1 Date of Issue: 18 June 2018
7.2.2 Test Status
The status is shown on the CMS screen. This contains a number indicating the current state of the CMS, according
to the table below. This allows a system to remotely monitor the CMS operation, if desired, allowing more specific
diagnostics
(B)
.
Value
Function
Comment
0
NOT READY
Unit is powering up or there is some problem
1
READY
Ready to start a test
(C)
2
TESTING
Test in progress
3
WAITING
Waiting between tests
(D)
128
FAULT OPTICAL
LED failure/ sensor blocked/ filled with air
129
FAULT FLOW LOW
Flow too low for reliable test
(E)
130
FAULT FLOW HIGH
Flow too high for reliable test
131
FAULT LOGGING
Fault with data logging
132
FAULT WATER SENSOR
Water sensor failure
Table 7A
(B) However the fault conditions are also indicated on the front panel LED, while ``No Result’’ in the case of a
fault is indicated using special result values as previously described.
(C) User has not set tests to occur automatically.
(D) User has set a non-zero test interval.
(E) Or fluid is totally clean (no particle counts). Flow alarm can be turned off by user if this is a problem, for
example cleaning rigs.
7.2.3 Other Faults
Unexpected results obtained from sample
Check that the microbore pressure hose has been fully connected at both the system and CMS ends.
NOTE: removing either hose from the CMS is not indicative of flow through the CMS as the hose end is
now at atmosphere. The pressure drop across the CMS should be verified, see section 6.1.2.1.
Confirm that the flow through the CMS is within the range of the unit.
High water / aeration levels
Remote Device dialogue not responding to buttons being pressed
Check that correct COM port has been selected in the Remote Device dialogue.
Check USB driver has been installed.
Disconnect power supply to CMS and then reconnect it.
If the CMS has been subjected to excessive contamination and a blockage is suspected, a flush with a suitable
solvent may clear the blockage.
The standard CMS is fitted with Viton seals, so Petroleum Ether or Iso Propyl Alcohol may be used for this
purpose, in conjunction with the RMF Systems Bottle Sampling Unit. See flushing guidelines on supplied product
USB.
DO NOT USE ACETONE

Table of Contents