28
Meter Maintenance and Troubleshooting
7
When to clean or disinfect the
meter:
Clean the meter to remove visible dirt
or other material.
Disinfect the meter between each
patient use.
What to clean or disinfect:
The area around slots and openings
The meter display
The entire meter surface
1
Make sure the meter is turned o.
Gently wipe the meter surface with a
soft cloth slightly dampened (wring
out any excess liquid) with one of
these solutions:
To clean the meter
Mild dishwashing liquid mixed with
water
To disinfect the meter
70 % isopropyl alcohol
Note: During disinfection, make sure
that the meter surface stays wet with
isopropyl alcohol for 2 minutes.
Additional wiping of the surface with
cloths dampened with isopropyl
alcohol may be necessary.
2
Dry the meter thoroughly with a soft
cloth.
Error Messages
w WARNING
Risk of a serious health incident
Using an error message to change
your therapy can lead to improper
therapy decisions.
Never make therapy decisions based
on an error message.
If you have any concerns or see any
other error message, contact Roche.
NOTE
The error code and Err will alternate
on the display for all coded error
messages (E-1 through E 14).
The meter will not turn on or the
display is blank.
Batteries are dead.
Insert new batteries.
Display is damaged. / Meter is
defective.
Contact Roche.
Extreme temperatures.
Move the meter to a location with a
more moderate temperature.