Control Tests
27
4
When to Perform a Control Test
Performing a control test lets you know the meter and test strips are working properly. You should
perform a control test when:
• you open a new test strip box.
• you left the test strip container open.
• you think the test strips are damaged.
• you want to check the meter and test strips.
• the test strips were stored in extreme temperatures, humidity, or both.
• you dropped the meter.
• your test result does not match how you feel.
• you want to check if you are performing the test correctly.
Chapter 4: Control Tests
About the Control Solution
• Use only Accu‑Chek Performa control solution.
• Close the control solution bottle tightly after use.
• Write the date you open the control solution bottle on the bottle label. The control solution must be
discarded 3months from the date the control solution bottle was opened (discard date) or on the useby
date on the bottle label, whichever comes first.
• Do not use control solution that is past the useby or discard date.
• Refer to the control solution package insert for control solution storage conditions.
• The meter automatically recognizes the dierence between the control solution and blood.
• The control results are not displayed in memory.
• The control solution can stain fabric. Remove stains by washing with soap and water.
79718_07255233050_01_EN.indb 27 5/22/2019 1:44:54 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Accu-Chek Performa Nano and is the answer not in the manual?

Roche Accu-Chek Performa Nano Specifications

General IconGeneral
BrandRoche
ModelAccu-Chek Performa Nano
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals