7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes for extended host interface 119
About block codes for extended HIF
The following table lists the modified and added block
codes. It also shows the request/response behavior for
host-instrument communication.
Block composition notation
Host (request) Instrument (response)
Block code Request Block code Request
02 Calibration Result
03 Control Result
04 Sample Result
05 Calibration Result with Lot Information
06 Control Result with Lot Information
07 Sample Result with Lot Information
20 Operation Setting 21 Operation Setting Response
25 Service Action Execution 26 Service Action Response
27 Service Action Result Request 28 Service Action Result Response
29 Failed Service Action Result Response
45 Test Status Request 46 Test Status Response
47 Failed Test Status Request
52 Calibration Status Request 54 Calibration Status Response
57 Failed Calibration Status Request
50 Cassette Status Request 55 Cassette Status Response
59 Failed Cassette Status Request
64 Automatic service action status
69 Service Request Response
75 Message Log Request 76 Message Log Response
77 Failed Message Log Request
80 Sample Order Status Request 81 Sample Order Status Response
85 Failed Order Status Request
90 System Status Request 91 System Status Response
94 Date and Time Request 95 Date and Time Data
96 Serial Number Request 97 Serial Number Response
y List of block codes
Line
code
Description Iteration
55 Test ID
1 to m
(1)
(1) Number of iterations in all lines
05 Access Date Time
16 CALidx
11 Reagent Lot
Information
02 Control ID
1 to n
(2)
(2) Number of iterations in lines 02 and 12
12 Control Lot
Information
y Notation of block codes