4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes 65
Block 05: Calibration Result with Lot Information
Block 05 is transmitted to the host if all of the following
conditions are met:
• After receiving a result request with result type
selector 11 or 18
• If an accepted and not yet transmitted calibration
result is available
• If the instrument is configured accordingly
Line
code
Description Iteration
(1)(2)
(1) m = 1 to 6. m is the number of standard rates in line 03. The
sequence of m iterations must correspond to the sequence of values
in line 03. In addition, the number of iterations m has to be equal to
the number of parameters (standard rates) in line 03.
(2) y = 0 or 1. The value of y depends on whether the antigen excess
(AGE) check is switched ON or OFF. The instrument always transmits
an AGE rate line regardless of this parameter. If the AGE check was
not required by the test, this line contains the values initialized by the
instrument.
55 Test ID
15 Access Date Time
16 Result Indexes
03 Standard Rates 1 to y+1
04 Calibrate Curve
13 Calibrator Lot Information
11 Reagent Lot Information
00 Result Data 1 to m
07 ABS Sample Check 0 to y
y Block 5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio HT and is the answer not in the manual?

Roche Cedex Bio HT Specifications

General IconGeneral
BrandRoche
ModelCedex Bio HT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals