8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface 147
Line Code 19: Calibration Information
Example
19_11/04/2016_09:00:12_07-3718-
6•_176121••••_30/05/2016_62159401••_68354•••••_???
???????_00/00/0000_??????????_00/00/0000_?????????
?_00/00/0000_R0••••••_0.0007479•_F•••••••••_5286.7
620•_??????????_??????????_??????????_??????????_?
?????????_??????????_••0<LF>
Description Format Range
Line Code 19
Calibration Date D1 dd/mm/yyyy
Calibration Time T1 hh:mm:ss
Calibrator ID S10
Calibrator Lot Number S10
Calibrator Lot Expiration Date D1 dd/mm/yyyy
Cassette Lot Number used for Calibration S10
Cassette Serial Number used for Calibration S10
Calibrator ISE Solution 1 Lot Number S10
Calibrator ISE Solution 1 Expiration Date D1 dd/mm/yyyy
Calibrator ISE Solution 2 Lot Number S10
Calibrator ISE Solution 2 Expiration Date D1 dd/mm/yyyy
Calibrator ISE Solution 3 Lot Number S10
Calibrator ISE Solution 3 Expiration Date D1 dd/mm/yyyy
Calibration Result Coefficient 1 Label S10
Calibration Result Coefficient 1 S10
Calibration Result Coefficient 2 Label S10
Calibration Result Coefficient 2 S10
Calibration Result Coefficient 3 Label S10
Calibration Result Coefficient 3 S10
Calibration Result Coefficient 4 Label S10
Calibration Result Coefficient 4 S10
Calibration Result Coefficient 5 Label S10
Calibration Result Coefficient 5 S10
Calibrator Flag I3 Detailed information in the table below
y Line Code 19: Calibration Information