7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
122 Result blocks
Block 04: Sample Results
Block 04 is transmitted to the host if all of the following
conditions are met:
• After receiving a result request block 09
• If an accepted and not yet transmitted sample result is
available
Line code Description Iteration
(1)(2)(3)(4)(5)
(1) n = 1. If result type selector 01 or 07 is used (single result)
(2) n = 1 to 50. If result type selector 03 or 04 is used (results collated
by order number)
(3) w = 1 to 141. The value of w can be different from 1 only if the result
was measured using an absorbance test (Line 06). There are 141
possible lines since each line is produced for every cycle (max 141).
(4) p = 0 to x. The value of x is the number of controls used for a single
test result. It is possible that no control was used because QC was
never performed.
(5) y = 1 to z. The value of z is the total number of reruns executed for a
single test result.
53 Order ID
56 Sample Name 0 to 1
55 Test ID
1 to n
00 Result data
14 Validation status
06/08/09 Raw Data 1 to w
17 Test Execution
18 Cassette Information
19 Calibration Data
21 Control Data 0 to p
27 Rerun results 0 to y
y Block 04