1 About communication
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
24 Error handling
Error handling
Whenever the host detects any transmission error, it has
to re-transmit the previous message keeping the same
sequence counter as before.
Whenever the instrument detects any transmission error,
it expects the host to re-transmit the previous message,
keeping the same sequence counter as before. If you
want to skip this retransmission, send a synchronization
request to the instrument, then resume the normal
communication.
Possible transmission errors:
Incorrect block check sum, i.e. a corrupted message
was received
Parity error was detected
General error message (block 99) was received
Time-out occurred
u Related topics
Synchronization and sequence counter use (26)