8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
170 List of line codes for extended HIF
Line Code 97: System status
Example
97_•5_0_0_0_1_•853_0_•••37.1_•••12.6_•••24.2<LF>
Description Format Range Range description
Line Code 97
System status I2 0 Sleeping
1 Initializing
2 Stopped
3 Service
4 Cover Open
5 Standby
6 Running
7 Fatal Error
8 Power Down
9 Auto Service
10 Power On
Cleaner status I1 0 Ok
1 Warning
2 Error
Cuvette status I1 0 Ok
1 Warning
2 Error
Water supply
status
I1 0 Ok
1 Warning
2 Error
Cuvette waste
status
I1 0 Ok
1 Warning
2 Error
Cuvette waste
count
I4 0..1000
Waste drain
status
I1 0 Ok
1 Warning
2 Error
Analyzer
temperature
F7 Temperature in degrees
Cassettes
temperature
F7 Temperature in degrees
Cleaner
temperature
F7 Temperature in degrees
y Line Code 97: System status

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio HT and is the answer not in the manual?

Roche Cedex Bio HT Specifications

General IconGeneral
BrandRoche
ModelCedex Bio HT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals