CoaguChek XS System User Manual for Self-Testing, version 10.0 USA
62
Error: Meter Ambient
Temperature
The meter is too cold or
too warm to measure
correctly.
Solution:
Turn the meter off and
allow it to stand for about
30 minutes at room
temperature between
59°F and 90°F (15°C and
32°C).
Notice
If the code chip of a liquid quality control solution
(number starting with “C_…”, not a regular
consumable for this meter) has inadvertently been
inserted, the “Meter Temperature” error may also be
displayed. In this case:
Remove the wrong code chip and the batteries.
Reinsert the batteries after no less than 5 minutes.
Set date and time correctly after inserting the
batteries (see the chapter Setting the Date and Time
starting on page 19 to set up date and time
correctly).