Operation Power Supply NGMO
4.00 / 07-2004 4-13
A/B11: Trigger Source
[int] ext
The trigger source function lets you select either an
internal trigger source (see A/B07 to A/B09) or use an
external trigger signal to start an analysis measurement.
The external trigger signal can be applied on the rear
panel.
A/B12: Trigger Slope
[pos] neg
The trigger slope specifies the slope (positive or negative
transition) used for signal triggering. A positive slope is a
low-to-high current or voltage transition, a negative slope
specifies a high-to-low transition.
A/B13: Trigger Count
(1-100)
The trigger count specifies the number of times the
analysis measurement is repeated to calculate the average
value for: Average, Peak, Min, High, Low and RMS
functions. Any value from 1 to 100 can be selected.
A/B14: Trigger Offset
(-5000-50000 points)
The trigger offset specifies a trigger delay for analysis
measurements in terms of a certain number of sampling
points. A positive “Trigger offset” specifies a delay, a
negative “Trigger offset” specifies an advance trigger (pre
trigger).
Any integer value between -5000 to 50000 can be selec-
ted.
A/B15: Sample length
(1-5000 points)
The sample length is the number of samples that are taken
for a measurement signal analysis.