EasyManua.ls Logo

Rohde & Schwarz NGMO1 - Page 34

Rohde & Schwarz NGMO1
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Power Supply NGMO
4.00 / 07-2004 4-13
A/B11: Trigger Source
[int] ext
The trigger source function lets you select either an
internal trigger source (see A/B07 to A/B09) or use an
external trigger signal to start an analysis measurement.
The external trigger signal can be applied on the rear
panel.
A/B12: Trigger Slope
[pos] neg
The trigger slope specifies the slope (positive or negative
transition) used for signal triggering. A positive slope is a
low-to-high current or voltage transition, a negative slope
specifies a high-to-low transition.
A/B13: Trigger Count
(1-100)
The trigger count specifies the number of times the
analysis measurement is repeated to calculate the average
value for: Average, Peak, Min, High, Low and RMS
functions. Any value from 1 to 100 can be selected.
A/B14: Trigger Offset
(-5000-50000 points)
The trigger offset specifies a trigger delay for analysis
measurements in terms of a certain number of sampling
points. A positive “Trigger offset” specifies a delay, a
negative “Trigger offset” specifies an advance trigger (pre
trigger).
Any integer value between -5000 to 50000 can be selec-
ted.
A/B15: Sample length
(1-5000 points)
The sample length is the number of samples that are taken
for a measurement signal analysis.