Power Supply NGMO IEEE Commands
5-30 4.00 / 07-2004
5.10 Remote calibration command summary
Com-
mand
Parameter Par. value Description Def. SCPI
:CAL :ENTer :PROT :CODe <NUM_VAL> 1...99999999 Unlocks
calibration code
2222
:CAL :CHANge :PROT :CODe <NUM_VAL> 1...99999999 Sets new
calibration code
:CAL [:A] Channel A /
Channel 1
:CAL[1] Channel A /
Channel 1
:CAL :B Channel B /
Channel 2
:CAL2 Channel B /
Channel 2
:STEP1 Sets output to
50mV
:STEP2 <NUM_VAL> 0-200 mV Reads Val from
external DVM
:STEP3 <NUM_VAL> 14.000-
16.000V
Reads Val from
external DVM
:STEP4 Turns on measure
voltage 5Volts
:STEP5 <NUM_VAL> 0.800 to
1.200 Ohms
Enters exact value
of 1 Ohm cal
resistor and do 5
Volts, 5A
Calibration
:STEP6 Turns on measure
Voltage 15 Volts
:STEP7 <NUM_VAL> 28.000 to
38.000 Ohms
Enters exact value
of 33 Ohms cal
resistor and do 15
Volts, 500 mA
Calibration
:STEP8 Turns on measure
Voltage 15 Volts
:STEP9 <NUM_VAL> 2800.0 to
3800.0 Ohms
Enters exact value
of 3 KOhms cal
resistor and do 15
Volts, 5mA
Calibration
:DATE <NUM_VAL>,
<NUM_VAL>,
<NUM_VAL>
YY, MM, DD Sets calibration
data
:SAVE Save calibration
data
:CAL :LOCK Locks calibration