Power Supply NGMO Operation
4-24 4.00 / 07-2004
sample value. That is important to protect against losing information
when observing signals over a long time period.
The sample interval can be set between 10 µs and 1 s in 10 µs steps.
At sample intervals greater than 10 µs the NGMO2 measure system
samples internally in 10 µs intervals and averages the taken samples to
build values in the selected time intervals.
Example:
With a selected sampling interval of 1.00 ms every sample value that is
stored in the measurement buffer is an average out of 100 samples.
Fig. 4-4 Sample values
The measurement buffer is able to store up to 5000 sample values
which can be read out separately to generate graphic representations
and which are used to build the specific pulse values as AVERage,
PEAK, MIN, HIGH, LOW and RMS.
By selecting a sample Intervall of 1 s and the maximum sample number
of 5000, the NGMO2 is able to observe a signal for a time period of mo-
re than 80 minutes without losing significant information about the sig-
nal course.