EasyManua.ls Logo

Rohde & Schwarz NGMO1 - SOURCE Command Summary

Rohde & Schwarz NGMO1
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IEEE Commands Power Supply NGMO
4.00 / 07-2004 5-17
5.5 SOURCE command summary
Com-
mand
Parameter Par. value Description Def. SCPI
:SOURce [:A] Channel A /
Channel 1
:SOURce
[1]
Channel A /
Channel 1
:SOURce :B Channel B /
Channel 2
:SOURce2 Channel B /
Channel 2
:VOLTage :PROTec-
tion
<CHAR_VAL> or
<NUM_VAL>
Sets VPT offset in
100mV steps
2.0
1.5 1.5 Volt
22 22 Volt
MIN 1.5 Volt
* DEFault 22 Volt
MAX 22 Volt
DISPlay Sets LCD to
VOLTAGE
PROTECTION
:VOLTage :PROTec-
tion?
<CHAR_VAL> Queries VPT
offset
Results:
0.0 <= NUM_VAL
<= 10.0
MIN 0.00
DEFault 2.0
MAX 10.0
:VOLTage :PROTec-
tion
:STATe? Queries VPT
state
Results:
0 -> not triggered
1 -> triggered
:VOLTage :MAXSet-
ting
<CHAR_VAL> or
<NUM_VAL>
Sets max. settable
output voltage
15
0.00 0 Volt
15.00 15 Volt
MIN 0 Volt
* DEFault 15 Volt
MAX 15 Volt
DISPlay Sets LCD to
Maximum output
voltage
:VOLTage :MAXSet-
ting?
<CHAR_VAL> Queries max.
settable output
voltage
Results:
0.000 <=
NUM_VAL <=
15.000