A/B20: READ xxx Value
(Result)
The “READ xxx Value” menu item (xxx = AVER PEAK MIN
HIGH LOW RMS) (see also A/B19) displays the result for
the selected analysis function. If there is no result available
(because there is no data that can be analyzed) only a line
is displayed.
A/B21: Maximum output
Voltage (0.000-15.000 V)
This menu item is an NGMO2 safety feature. Setting a
voltage value below 15.000 volts limits the maximum
settable output voltage in the main display to the voltage
given here. A connected device is therefore protected from
voltages above the maximum allowed ratings.
A/B22: Maximum output
Current (0.000-5.000 A)
This menu item is an NGMO2 safety feature. Setting a
current value below 5.000 Amperes limits the maximum
settable output current limit in the main display to the
current limit given here. A connected device is therefore
protected from currents above the maximum allowed
ratings.
A/B23: Beep on Current
limit: [on] off
When “Beep on Current limit” is set on the NGMO2, there
is a short beep whenever the NGMO2 enters the current
limit state.
A/B24: Common Output on
off: [on] off
When “Common Output on off” is set to OFF, the key
“OUTP. on/off” switches only the output channel that is
currently displayed or selected via remote command ON or
OFF. If “Common Output on/off” is set to ON, the “OUTP.
on/off” key acts simultaneously on both output channels. A
remote command also acts on both channels.