IEEE Commands Power Supply NGMO
4.00 / 07-2004 5-7
Com-
mand
Parameter Par. value Description Def. SCPI
:BAND-
width
<CHAR_VAL> Selects slow or fast
regulation speed
HIGH
9
HIGH Bandwidth high
LOW Bandwidth low
MIN Bandwidth high
* DEFault Bandwidth high
MAX Bandwidth low
DISPlay Sets LCD to
OUTPUT
BANDWIDTH
:BAND-
width?
<CHAR_VAL> Queries bandwidth
settings
Results:
LOW
HIGH
MIN HIGH
DEFault HIGH
MAX LOW
:IMPe-
dance
<CHAR_VAL> or
<NUM_VAL>
Specifies the
output impedance
to apply. 0 Ohms to
1Ohms in 10
mOhm steps
0
Ohms
0.00 0 Ohm
1.00 1 Ohm
MIN Impedance 0 Ohm
* DEFault Impedance 0 Ohm
MAX Impedance 1 Ohm
DISPlay Sets LCD to
OUTPUT
IMPEDANCE
:IMPe-
dance?
<CHAR_VAL> Queries impedance
settings
Results:
0.00 <= NUM_VAL
<= 1.00
MIN 0.00
DEFault 0.00
MAX 1.00
:OUTPut [:A]
:RELay1 Relay 1
:RELay2 Relay 2
:RELay3 Relay 3
:RELay4 Relay 4
<CHAR_VAL> Closes or opens
relay control circuit
ZERO
or OFF
ZERo Opens
ONE Closes
OFF Opens