Remote control commands
R&S
®
RTE
1966User Manual 1326.1032.02 ─ 20
POWer:ONOFf:REPort:ADD......................................................................................... 1966
POWer:ONOFf:RESult:TOFF?...................................................................................... 1967
POWer:ONOFf:RESult:TON?........................................................................................ 1967
POWer:ONOFf:TIME....................................................................................................1967
POWer:ONOFf:TYPE................................................................................................... 1967
POWer:ONOFf:ATOFf <ACTrigLevOff>
POWer:ONOFf:ATON <ACTrigLevOn>
Triggers the beginning of the measurements at the moment the AC input voltage rea-
ches the set value.
Parameters:
<ACTrigLevOn> Range: -1E+6 to 1E+6
Increment: 1E-3
*RST: 10
Default unit: V
POWer:ONOFf:DTOFf <ACTrigLevOff>
POWer:ONOFf:DTON <ACTrigLevOn>
POWer:ONOFf:DSOFf <StateLevelOff>
POWer:ONOFf:DSON <StateLevelOn>
Sets the percentage of the steady state level of the DC output that has to be reached.
Parameters:
<StateLevelOn> Range: 0 to 100
Increment: 1
*RST: 90
Default unit: %
POWer:ONOFf:EXECute
Starts the turn on/off measurement.
Usage:
Event
POWer:ONOFf:INPut <InputType>
Sets the input type.
Parameters:
<InputType> AC | DC
*RST: AC
POWer:ONOFf:REPort:ADD
Adds the result to the report list.
Usage:
Event
Power analysis (option R&S RTE-K31)