Mixed signal option (MSO, R&S RTE-B1)
R&S
®
RTE
918User Manual 1326.1032.02 ─ 20
"Width"
Sets a pulse width as timing condition, see "Width settings"
on page 918. The pulse starts when the pattern comes true, and the
trigger event occurs when the pattern comes false during the speci-
fied time limit.
Using this mode, you can, for example, trigger exclusively on unsta-
ble conditions - if the pattern is present for less than a specified time.
Remote command:
TRIGger<m>:PARallel:PATTern:MODE on page 1918
Timeout settings
The timeout settings "Range" and "Time" appear if the timing mode is set to "Timeout".
Range ← Timeout settings
Sets the state condition:
"Stays high"
The pattern stays true for the specified time.
"Stays low"
The pattern stays false for the specified time.
"High or low"
The pattern remains unchanged for the specified time.
Remote command:
TRIGger<m>:PARallel:PATTern:TIMeout:MODE on page 1918
Time ← Timeout settings
Defines the time limit for the timeout at which the instrument triggers.
Remote command:
TRIGger<m>:PARallel:PATTern:TIMeout[:TIME] on page 1919
Width settings
The width settings "Range", "Width" and "±Delta" appear if the timing mode is set to
"Width".
Range ← Width settings
Selects how the range of a pulse width is defined:
"Within"
Triggers when the pattern comes false inside a given time range. The
time limit is defined by Width
±
Delta.
"Outside"
Triggers when the pattern comes false before or after the given time
range. The time limit definition is the same as for "Within" range.
"Shorter"
Triggers when the pattern comes false before the given "Width" has
expired.
"Longer"
Triggers when the pattern comes false after the given "Width" has
expired.
Remote command:
TRIGger<m>:PARallel:PATTern:WIDTh:RANGe on page 1919
Width ← Width settings
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
time limit, respectively.
Trigger