Performing Measurements
R&S
®
ZNB/ZNBT
68Getting Started 1316.0062.02 ─ 26
Tip: For a R&S ZNBT with more than 4 ports, the graphical port representation is
replaced by a generic port list. The selection logic is unchanged.
4. Select "Next" to proceed to the next page of the "Calibration Setting" wizard.
5. Select the test port connector type and gender (here: N 50 Ω, female, correspond-
ing to a male Through standard) and the calibration kit (here: R&S ZV-Z121), then
click "Start".
6. The "Calibration" dock widget indicates the standard measurements that make up
a "Trans Norm" calibration.
Transmission S-Parameter Measurement