14
Reference Manual
00809-0100-4378, Rev CE
Appendix A: Reference Data
February 2014
Reference Data
A.2 LCD display specifications
A.2.1 Functional specifications
Input signal
4–20 mA dc
Display
4 mA point limits
–999 to 1000
20 mA point limits
-999 to 9999
The sum of the 4 mA point and span must not exceed 9999. Adjustments are made using
non-interactive zero and span buttons.
Display options
Standard display response is linear with mA input. Optional square root or filtered response may
be selected.
Overload limitations
666 mA, maximum
Temperature limits
Storage
–40 to 85 °C (–40 to 185 °F)
Operating
–40 to 70 °C (–40 to 158 °F)
(1)
Humidity limitation
0 to 95% non-condensing relative humidity
Update period
750 ms
(1) For temperatures below -20 °C or above 60 °C the LCD may not be readable, but the loop will remain intact and the LCD will not
be damaged.