3.9.5 CPR (mechanical or electrical)
Note: The side rails of the bed should be in raise position before triggering the CPR
function. Trigger CPR as follows:
- (Mechanical) Pull the CPR handle under the thighs section platform as shown below
to lower the back section only.
- (Electrical) Press on the button on the nurse control (refer to section 3.3.2) to get the
platform to move automatically to a completely horizontal position.
Caution! The back section will automatically fall to horizontal position by holding the
CPR handle or the press button (electrical CPR will also lower the thighs section). Ensure
that no member of the user exceeds the mattress and no objects or equipment is not in the
head section of the bed platform, to avoid the risk of injury and / or breakage.
Note :
The total lock function does not affect the CPR function.
A second handle can be installed in option
Caution! If electrical CPR function button is released within 4 seconds the action will
stop. Otherwise, if the button is held for more than 4 seconds and then released, the action
will automatically continue. Press any key on the nurse control panel or patient control
panel to stop and cancel the automatic movement.