EasyManua.ls Logo

Rotex A1 BO - Burner Maintenance and Adjustment

Rotex A1 BO
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 x Oil burner
44
FA ROTEX A1 BO - 10/2007
Tools required: Flue gas analysis unit for determining the emission values for NO
x
and CO in the flue gas.
Connect the flue gas analysis unit.
Adjust the position of the recirculation gap with adjusting screw (for setting, see tab.6-1).
Shut down and wait for about 5 minutes.
Restart the burner.
Î Burner does not start or is late: Reduce the width of the recirculation gap (lower scale value).
Check emission values for NO
x
and CO (observe country-specific regulations).
6.3.6 Check and set electrode distance and the distance between oil nozzles and air nozzles
Checking and setting electrodes distance
Tools required: Feeler gauge "Blue Flame Meter" (fixed on the holder for the service position);
Put burner in service position (see chapter 6.4).
Check electrode distance and position with feeler gauge "Blue Flame Meter".
If necessary, readjust electrodes by bending them.
CAUTION!
If the recirculation gap is too small or closed, it will cause a sudden temperature rise in the mixing device and
can damage the mixing device.
a Mixing unit
b Ignition electrodes
c Fixing screw
d Light tube
e Oil nozzle
f Air nozzle
g Nozzle rod
X1 Electrode position BL 15, BL 20
and BL 27
X2 Electrode position BL 35
Y Distance oil nozzle – air nozzle
Fig. 6-8 Check electrodes setting and distance oil nozzles - air nozzles
If the ignition electrodes are worn out, they must be replaced.
Notes on replacing the ignition electrodes are given in section 8.2.8.

Table of Contents

Related product manuals