Set the output voltage at 1V, adjust the output voltage zero point to make the
multimeter display close to 1V, and then set the output voltage at 30V, adjust the
output voltage proportional value to make the multimeter display close to 30V. In the
same way you can set 0.1A and 3A output current to calibrate the zero point and
proportional value of the output current.
Set the output voltage at 1V and calibrate the actual output voltage zero point to
make the actual output voltage displayed on RK6006 close to the value on
multimeter. You can set 30V and calibrate the proportional value of actual output
voltage. In the same way you can set 0.1A and 3A to calibrate the zero point and
proportional value of the actual output current. (This section does not provide
technical support. If you do not understand, please check the related information).
4.3.3 Advanced Function
You can set the output voltage and current by chart in the advanced function page,
you can set every step between 1 and 9999 seconds, you can set 200 steps max, it
can output automatically or manually. You cannot choose other operation page when
it performs programming output or other operations, you can only switch other page
when it ends.